Obiat Pty Ltd distributors of Electronic and Professional Scientific Test Equipment.

 

Sydney Australia

. . . . keeping your world up and running

Navigation: Fluke 110 Series Index > Model 117 DMM > Specifications > Model/Accessories
   

   
 
 
 
Maximum voltage 600 volts (between any terminal and earth ground)
Surge protection 6 kV peak per IEC 61010-1 600V CAT III, Pollution Degree 2
Fuse for 'A' input 11A, 1000V Fast Fuse (Fluke PN 803293)
Display Digital: 6000 counts, updates 4 per second
Bar graph 33 segments, updates 32 per second
Operating Temperature -10°C to + 50°C
Storage Temperature -40°C to +60°C
Battery type 9 volt Alkaline, NEDA 1604A/IEC 6LR61
Battery life 400 hours typical, without backlight
 
 
 
DC millivolts
  Range:
  Resolution:
  Accuracy:

 

600.0mV

0.1mV

±([%of reading] + [counts]):0.5% +2

DC volts
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Accuracy:

 

6.000V/0.001V

60.00V/0.01V

600.00V/0.1V

±([%of reading] + [counts]):0.5% +2

Auto volts  
  Range:
  Resolution:
  Accuracy:
   

 

600.0V

0.1V

2.0% + 3 (dc, 45Hz to 500 Hz)

4.0% + 3 (500Hz to 1kHz)

AC millivolts1 true-rms
  Range:
  Resolution:
  Accuracy:
   

 

600.0mV

0.1mV

1.0% + 3 (dc, 45Hz to 500 Hz)

2.0% + 3 (500Hz to 1kHz)

AC volts1 true-rms
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Accuracy:
   

 

6.000V/0.001V

60.00V/0.01V

600.00V/0.1V

1.0% + 3 (dc, 45Hz to 500 Hz)

2.0% + 3 (500Hz to 1kHz)

Continuity  
  Range:
  Resolution:
  Accuracy:
   

 

600 ohms

1 ohm

Beeper on <20Ω, off >250Ω; detects opens or shorts of

500µs or longer

Ohms
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Accuracy:
  Range/Resolution:
  Accuracy:
 

600.0Ω/0.1Ω

6.000kΩ/0.001kΩ

60.00kΩ/0.01kΩ

600kΩ/0.1kΩ

6.000MΩ/0.001MΩ

0.9% + 1

40.00MΩ/0.01MΩ

1.5% + 2

Diode Test
  Range:
  Resolution:
  Accuracy:
 

2.000V

0.001V

0.9% + 2

Capacitance
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Accuracy:
  Range/Resolution:
  Accuracy:
 

1000nF/1nF

10.00µF/0.01µF

100.0µF/0.1µF

9999µF/1µF

100µF to 1000µF

1.9% +2

> 1000µF

5% + 20%

LoZ Capacitance
  Range:
  Accuracy:
 

1nF to 500µF

10% + 2 typical

AC amps true-rms
(45Hz to 500Hz)
  Range/Resolution:
  Range/Resolution:
  Accuracy:
   
 
 

6.000A/0.001A

10.00A/0.01A

1.5% + 3

20A continuous overload for 30 seconds maximum

DC amps
  Range/Resolution:
  Range/Resolution:
  Accuracy:
   
 

6.000A/0.001A

10.00A/0.01A

1.0% + 3

20A continuous overload for 30 seconds maximum

Hz (V or A input)2
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Range/Resolution:
  Accuracy:
 

99.99Hz/0.01Hz

999.99Hz/0.1Hz

9.999Hz/0.001Hz

50.00Hz/0.01Hz

0.1% + 2

 
Notes:  

1)

All AC voltage ranges except Auto -V/L0-Z are specified from 1% to 100% of range.

 

Auto -V/L0-Z is specified from 0.0V.

2)

Frequency is AC coupled, 5Hz to 50kHz for AC voltage.

 
 
 
Size 167mm x 84mm x 46mm (6.57" x 3.31" x 1.82")
Weight 550g
Warranty 3 years


 

AEMC | AMPROBE | CIE | FLUKE | HT ITALIA | HODEN | PROVA | POMONA | RAYTECH |  SILVERTRONIC
Home | PRODUCTS | SEARCH
Obiat Pty. Ltd.
A.C.N. 003 166 952      A.B.N. 23 003 166 952
Unit 9A/8 Avenue of the Americas, Newington NSW 2127 Australia
P.O. Box 6901 Silverwater  NSW Australia 1811

Phone: 61-2-9748- 8699

Fax: 61-2-9748-8544

E-mail:
 

 information@obiat.com.au
sales@obiat.com.au

Copyright ©                                             Last Update: Wednesday, 13 April 2011